app_link.jpg imurmur_link.jpg
OOPS. Your Flash player is missing or outdated.Click here to update your player so you can see this content.
  Home arrow Support and FAQs arrow Manuals & Documents arrow Online User Manual
Thursday, 09 September 2010
 
ds32a User Manual
Article Index
ds32a User Manual
Page 2
Page 3
Page 4
Page 5
Page 6
Page 7
Page 8
Page 9
Page 10
Page 11
Page 12
Page 13
Page 14
Page 15
Page 16
Page 17
Page 18
Page 19
Page 20
Page 21
Page 22
Page 23
Page 24
Page 25
Page 26


Listening with the ds32a

The ds32a Probe uses Electromagnetic Diaphragm (EmD) technology which has some important characteristics that should be understood to achieve optimal performance. This background information is provided simply to enrich your understanding of the ds32a, so that you can obtain maximum benefit from its features. Read this section while experimenting with your ds32a, and this background information will become obvious and self-evident. If the explanation below seems complicated, rest assured using the ds32a stethoscope is very simple.

Diaphragm Pressure and Frequency Response

The EmD Diaphragm is pressure-modulated i.e. pressure controls the mechanical and electrical parameters of the EmD sensing system thereby changing the frequency response and sensitivity. The user can significantly alter the sound by adjusting diaphragm pressure. Learn how to use this to advantage during patient examination - it is a powerful, tactile way to control the sound you hear, and adjust the sensitivity for each patient examination. With a little practice, you will soon be able to control the way you listen, to elicit maximum benefit.

Apply the Probe to the chest, make subtle changes in diaphragm pressure, and listen to the effects. The following control is possible:

- Light Pressure - This is usually optimal for most listening. Apply sufficient pressure to get good skin contact and a clear signal, but do not push too hard.

- Increasing Pressure - Gently increasing pressure will increase the sensitivity of the EmD sensor, making sounds louder. This is useful for listening to obese patients, listening for faint bruits, etc. However, the combination of loud body sounds (such as listening to the heart at the apex on a thin patient) and increased pressure (high sensitivity) can produce distortion due to excessive signal levels in the sensor. It is therefore not necessary to increase sensitivity when sounds are already loud and clear. Apply diaphragm pressure according to the patient and the probe position. The purpose is to provide a tool so powerful and flexible that its characteristics are suitable for a wide range of patients and body sounds.

- Significant Pressure - With sufficiently high pressure, the diaphragm makes contact with the sensing plate behind it, and diaphragm vibration is restricted. At that point, low frequencies are attenuated, and you will hear breath sounds over heart sounds. This can be a useful mode for reducing heart sounds to listen to other sounds, however care should be taken not to work in this mode when heart sounds are the sounds of interest. If low frequencies suddenly become attenuated due to excess pressure, simply back off diaphragm pressure slightly.

Within a short time, it will become intuitive to adjust the Probe/diaphragm pressure for a particular examination and you will enjoy the control that you have over sound characteristics.

hex142.jpgHolding the Probe - steady pressure

It is helpful to hold the Probe with a consistent pressure, so that sound can be controlled as described above. One method is to hold the Probe so that the fingers also touch the patient chest wall (such as the left edge of the thumb and the fingertips in the figure.) The Probe and chest wall then move in unison, with a consistent, light diaphragm pressure. This is merely one grip - you will find your own personal preferences for optimally using your ds32a.

foot_note




 


© 2010 Thinklabs Medical LLC