• Instructions - The ds32a Stethoscope is designed to be very easy-to-use. However,
taking the time to read the User’s Manual will greatly facilitate the
effective and optimal use of the ds32a. These instructions provide
valuable tips that are well worth knowing, to get the most out of the
ds32a.
• Sound Levels - Adjust the Volume to a comfortable level. The ds32a
Stethoscope has powerful amplification. Maximum volume is not necessary in most
situations. Experiment to identify optimal settings and personal
preferences.
• ANR2 Ambient Noise Rejection - The ds32a Stethoscope has second generation
Ambient Noise Rejection (ANR2), enhancing the noise rejection
capability of EmD technology. The noise rejection function can be
activated or deactivated, providing you with complete control over
signal processing functions. Experiment with each setting to select the
best for each situation.
• Diaphragm Pressure and skin contact - The ds32a Probe uses
Electromagnetic Diaphragm (EmD) technology which is pressure-sensitive.
Sound characteristics change as the Probe is applied with greater pressure to
the patient. Simply apply the Probe to the patient as with a
conventional stethoscope, and make subtle changes in pressure,
listening for desired sound. Lighter pressure is usually optimal.
Increase pressure gradually to increase sensitivity and volume. With
significantly increased pressure, low frequencies will suddenly be
attenuated and lung sounds will be audible over heart sounds. Apply
steady diaphragm pressure for best results. The ds32a’s EmD diaphragm
operates best when in direct skin contact with the patient.
• Maintaining Performance - Your ds32a Stethoscope is a robust instrument designed
to provide continued peak performance. If your ds32a is not meeting
your highest expectations, please contact us so that we can help you
obtain and maintain the performance level for which it is designed.